- Laboratory Heating Equipments[1]
- Semiconductors[4]
- Other Electrical Equipment[2]
- Drying Equipment[10]
- Testing Equipment[8]
- Electronics Production Machinery[4]
- Packaging Boxes[1]
- Other Advertising Equipment[1]
- Other Consumer Electronics[10]
- Lab Drying Equipment[7]
- Electronic & Instrument Enclosures[2]
- Packaging Machinery[6]
- Other Metal Processing Machinery[5]
- Laboratory Furniture[6]
- Other Lab Supplies[4]
- Safes[10]
- Filing Cabinets[9]
- Motorcycle Ramps[9]
- Tool Cabinet[1]
- Other Vehicle Tools[1]
- Other Electronic Components[7]
- Other Service Equipment[1]
- Display Racks[1]
- Metal Processing Machinery Parts[1]
- Mailboxes[1]
- Contact Person : Ms. bo Kitty
- Company Name : Changshu Tongrun Electronic Co., Ltd.
- Tel : 86-512-52847596
- Fax : 86-512-52173876
- Address : Jiangsu,CHANGSHU,48Nijiaba Hongqiao Yushan zone Changshu city
- Country/Region : China
- Zip : 215500
Laboratory Vacuum Tube Furnace
Specifications: | |
Furnace structure | Double layer steel case with mullite liner without water cooling or air cooling |
Heating element | 8 pcs 1Kw Halogen light tube, ( Dia. =10mm, L=300 mm, Heated Length =200 mm)Standard working life : 2000 hrs. ( depended on heating rate ) |
Heating zone | 12" length with 4" constant temperature zone within +/-5ºC uniformity |
Quart Tube & Sample Holder | Quart Tube Size: 4.33" O.D x 4.05" I.D x 16.2" Length3" diameter Sample holder is inserted to sliding flange, which is made of Aluminum Nitride (AIN).+\- 5°C temperature uniformityOne 3" diameter AlN wafer is placed in the center of holder for annealing sample up to 3" diaAlN has very high thermal conductivity, which provides the best temperature uniformity in the sample areaThe sample holder is removable from flange and use the RTP for other purpose |
Vacuum Flange | Made of stainless steel with double hi-temp. silicone O-rings, KF25 Vacuum Port and Needle Valves.Slide-out flange provides an easier way of loading, unloading the samples and other specific process needed. |
Water Cooling Jacket | Cooling water is requiredPlease make sure the flow rate of cooling water is ≥ 10L/min, water temp. <25ºC; pressure > 25 PSIWe suggest you using recirculating water chiller as cooling system (please order separately at the related product below) to save water. Warning: This RTP furnace's max holding time shall be less than 600 seconds at 1000ºC ~ 1100ºC, otherwise O-ring may get melting |
Flow-meter | One flow meter is installed on the front panel to adjust gas flow from 16 - 160 ml/min |
Vacuum gauge | Digital vacuum gauge with a measurement range from 1 mtorr~800 torr is installed with the furnace. |
Vacuum pressure & Gas Fitting | Pressure can reach 50 mtorr by using mechanical vacuum pump with KF25 adapter.Pressure can reach 10-4 torr by molecular pump with KF25 adapter. |
Temperature control | PID automatic control via SCR (Silicon Controlled Rectifier) power control, e.g. current limiting phase angle fired the resistor. |
Thermal Couple | S type, the head of the thermocouple touches the Aluminum Nitride sample holder from underneath |
Programmable Temperature controller | Built in protection from overheated and broken thermal couple 30 programmable segments Auto tune functionTemperature control accuracy: +/- 1ºC Built in RS485 port for PC control Software is included. (Computer is not included)One 15" screen laptop with operation software installed is available as option at $900 extra cost. |
Max. Heating Rate | 50ºC/sec . |
Max. Cooling Rate | 60ºC/min (Under vacuum: 200 mtorr), 117ºC/min (Under atmosphere pressure) Optional: In order to achieve faster cooling rate < 10oC/second, you may choose RTP furnace with sliding rail. |
Working Temperature | 1100ºC Max. |
Power Input | 208-240 AC, 50/60 Hz single phase, 9 KW Max. (> 60A breaker required) |
Dimension | Flange closed: 760 X 330 X 530 mm (30"W x 13"L x 21" H); Flange opened: 1200 X 330 X 530 mm (47"W x 13"L x 21" H) |
Net weight | 45 kg (100 lbs) |
Warranty | One year limited warranty for electrical parts. (Consumable parts such as processing tubes, o-rings and halogen lamp are not covered by the warranty, please order the replacement at related products below.) |
Certificate | CE certified |
Optional | You may combine the RTP furnace with high vacuum station or gas livery system as vacuum furnace or CVD system. |
Warning | Warning: The tube furnaces with quartz tube are designed for using under vacuum and low pressure < 0.12 atm (absolute pressure) |